24-Hour Service Hotline
13475428416 15954827913
LFY-401A Static Voltage Decay Tester
The main purpose
Under
laboratory conditions, this instrument is suitable for determining the
electrostatic properties of fibers, yarns, fabrics, carpets, decorative fabrics
and other fabrics or various sheet-like products. The instrument host consists
of a corona discharge device and a probe detector. Using a given high-voltage
electric field, the fabric is discharged for a fixed time to induce static
electricity in the fabric, so as to detect the amount of static electricity,
the half-life of static voltage decay, and the residual amount of static
electricity to display the electrostatic performance of the tested fabric.
Instrument
characteristics
Whether the
fabric is seriously charged or not is determined on the one hand by the level
of static voltage after high voltage discharge, and on the other hand by the
retention time of this static voltage. The electrostatic escape rate on the
fabric is expressed by the electrostatic half-life, which is an important
evaluation index for testing the electrostatic performance of the fabric.
The instrument
consists of two parts: the main engine and the automatic control box, which can
be tested by the timing method and the constant pressure method respectively.
The instrument uses a digital voltmeter to directly display the instantaneous
peak value and half-life value of the fabric during high-voltage discharge.
Automatic shutdown of high voltage, automatic shutdown of the motor, easy to
operate.
Technical
indicators
1. The
measurement range of electrostatic voltage value is -10KV~0, and the accuracy
is ±5%
2. The
half-life time limit range is 0~99.99s/m/h
3. The diameter
of the turntable is (200±4) mm, and the speed is 1500 rpm
4. The distance
from the center of the turntable to the center of the sample and the center of
the electrode is (72±2) mm
5. Diameter of
sensing electrode plate (28±0.5) mm
6. The
discharge time is 0 to 99.99s (the experiment is scheduled to be 30 seconds ±
0.1 seconds)
7. Discharge
distance between needle electrode and sample:
8. Measuring
distance between the sensing electrode and the sample:
9. Sample test
size:
10. The
instrument has been calibrated for static voltage measurement before it is
issued.
11. Check as
required before use, the oil cylinder in the main engine must be full of oil
(transformer oil)
12. Power
supply: 220V 50Hz 1.5kW.
Applicable
Standards
GB/T
12703.1-2021 Textiles- -Test methods for electrostatic propensity-
Part 1: Test
method using corona charging
FZ/T 64082-2021 Scrim for filter media
Address: 195 Shandong Road, Qingdao Postcode: 266032
Tel: 0532-85614755 (Sales switchboard) 0532-85635588 (sales manager)
Fax: 0532-85653787 He Hongxia (Senior Engineer) Deputy Director 15954827913 QQ: 179584685
After-sales service tel: 0532-85624778 15192792891
魯ICP備10203291號-10 魯公網安備 37020302370559號 Support: LEIXUN